YS-JL929 Liquid Bacteria Retention Tester for Pharmaceutical Filter Membranes and Filters
Buscar

YS-JL929 Liquid Bacteria Retention Tester for Pharmaceutical Filter Membranes and Filters

【Technical Parameters】

1. Challenge Pressure System

1.1 Three-channel pressure control system: independently controlled and independently detects output pressure;1.2 Pressure Control Range: 10~400kPa;1.3 Pressure Display Accuracy: ±0.5% F.S. (Full Scale);1.4 At least 3 units of 1-gallon pressure vessels, used for containing challenge bacterial solution, wetting solution, and cleaning alcohol respectively;1.5 Pressure control mode: optional direct pressurization or gradual pressurization; the amplitude and frequency of gradual pressurization are configurable;

2. Sample Challenge System

2.1 Sample Challenge Assemblies: At least 12 sets of sample challenge assemblies, compatible with clamping samples of different sizes;2.2 Test Piping: Entire piping system can be autoclaved (high-temperature sterilizable);2.3 At least 6 channels of adjustable all-metal flow control valves;

3. Test Environment

3.1 Biosafety Cabinet: Complies with the requirements of YY 0569-2011 "Class II Biosafety Cabinets";3.2 Average Downward Air Velocity: 0.33±0.025m/s;3.3 Average Inlet Air Velocity: 0.53±0.025m/s;

【Configuration List】

1.Stainless steel pressure vessels (1-gallon): 3 units;

2.Aluminum busbar: 1 piece;

3.Test chambers: 12 units;

4.Air pipe: 3 meters;

5.Test chamber brackets: 2 pieces;

6.Air compressor: 1 unit;

7.Ultraviolet (UV) lamp tube: 1 piece;

8.Test chamber brackets: 2 pieces; (Note: Duplicate item retained as per original list; suggest verifying if it's a typo in the source text)

9.Biosafety cabinet: 1 unit.


Productos relacionados